Required Process Tolerance /

Weighing Accuracy 1%

Model

Maximum

Capacity [g]

Readability

[mg]

Repeatability at full load [mg]

Linearity deviation sd [mg]

Minimum Weight According to USP [mg]

Incorporated Calibration Weight

Typical Minimum Weight1 [g]

WMS

104C

120

0.1

0.12

0.25

0.16

Yes

0.016

WMS

204

220

0.1

0.20

0.40

0.24

No

0.024

WMS

404C

410

0.1

0.10

0.40

0.16

Yes

0.016

WMS

403

410

1.0

1.00

2.00

1.00

No

0.100

WMS

1203C

1220

1.0

1.00

3.00

1.60

Yes

0.160

WMS

803

820

1.0

1.00

3.00

1.60

No

0.160

WMS

6002C

6200

10.0

10.00

30.00

12.00

Yes

1.200

WMS

4002

4200

10.0

10.00

30.00

16.00

No

1.600

Back to the Select Site

Remarks: 1) The minimum weight achievable depends on the settings of the weighing device, the tare vessel as well as the environment. The minimum weight of your device in situ can therefore be smaller or larger than the typical values published, for which METTLER TOLEDO takes no responsibility. The determination of the minimum weight on site is documented in GWP® Verification.

Disclaimer: "These general recommendations are for information purposes only and are not binding in any way. To ensure weighing accuracy, it is necessary to conduct calibration and test procedures at the location of the weighing device in a regular and correct manner. This document is offered as a recommendation for a scale or scale according to user requirements. The Recommendation is based on published specifications of METTLER TOLEDO weighing equipment. Other factors, which might also have an influence on the performance of the equipment, such as type of installation, environment, tare containers, experience of operators, statistical variations, etc., have not been taken into account. Therefore the information given is to be considered as a recommendation. The final responsibility is with the user of the equipment. This recommendation does not extend our warranty in any way.”