Do you combine calibration with an external test weight?
Test and adjustment can be done by positioning a certified weight onto the weighing pan if the weigh module is accessible inside the machine.
It is recommended to combine tests with internal and external weights. The test with the internal weight is used to verify accuracy any time needed. The external test weight is used for traceability and to verify the internal test weight.